4-2 HOW TO AVOID MALFUNCTIONS
Specimen preparation:
When preparing specimens, be sure that a
suitable embedding medium, fixation,
dehydration and infiltration time are chosen.
Specimen temperature:
Sectioning is carried out at ambient
temperature (excluding frozen sections).
If the temperature is too high, the paraffin
softens. Therefore, avoid heating paraffin
specimens by direct exposure to sunlight or
other near sources of heat.
Tightening the clamping screws:
Tighten all clamping screws and clamping
levers on the knife carrier, specimen holder
and specimen orientation.
Selection of the knife:
Carefully select the required knife material and
profile.
Adjustment of the knife:
Take care to adjust the proper clearance angle
of the knife: 5 - 7° for glass knives. For
diamond knives ask the knife manufacturer.
Select a clearance angle adjustment of 5 - 15°
according to the facet angle. Typically, adjust
an angle of 10 - 12 °.
Take care to adjust knife height.
Cutting speed:
Take care to select proper cutting speed.
General Rule: The harder the material, the
slower the cutting speed!
Trimming:
Take care in bringing knife and specimen
together.