Moving the knife carrier on the console:
Loosen the clamping lever (fig. 44.5) on the left
side of the knife carrier to move the carrier
forwards and backwards on the dovetail guide.
This allows a rough adjustment of knife and
specimen.
Protection against injury:
The knife carrier is equipped with two knife
guards (fig. 44.7) which can be moved
sideways.
These knife guards should be pushed together
in the middle while adjusting knife or specimen.
This reduces the danger of injury considerably.
Central clamping plate:
The knife is clamped and stabilized in the
cutting zone by the central clamping plate (fig.
44.8) - exactly where the highest cutting forces
are applied.
Two types of clamping plates are available for the
knives:
clamping plate c for c-knives,
clamping plate d for d-knives.
Fig. 45 shows schematically the angles on the
cutting edge profiles of c- and d-knives.
Fig. 45