Moving the knife carrier on the console:
• Loosen the clamping lever (fig. 44.5) on the left
side of the knife carrier to move the carrier
forwards and backwards on the dovetail guide.
• This allows a rough adjustment of knife and
specimen.
Protection against injury:
• The knife carrier is equipped with two knife
guards (fig. 44.7) which can be moved
sideways.
• These knife guards should be pushed together
in the middle while adjusting knife or specimen.
• This reduces the danger of injury considerably.
Central clamping plate:
• The knife is clamped and stabilized in the
cutting zone by the central clamping plate (fig.
44.8) - exactly where the highest cutting forces
are applied.
Two types of clamping plates are available for the
knives:
clamping plate c for c-knives,
clamping plate d for d-knives.
Fig. 45 shows schematically the angles on the
cutting edge profiles of c- and d-knives.
Fig. 45