3-11-3 KNIFE CARRIER C
To insert the knife, the clamping screws (fig.
44.3) must be unscrewed slightly so the knife
can be pushed in from the side.
The height of the knife is adjusted with the two
knurled nuts (fig. 44.6) and the bar (fig. 44.7).
If the cutting zone of the knife cannot be used
anymore, it can be moved over its entire length
to the left and right side by loosening the
clamping screws.
This allows an optimal use of the entire knife
edge.
Caution:
When clamping the knife, please
tighten the two clamping screws (fig.
44.3) simultaneously.
Clearance angle adjustment:
The clearance angle between cutting edge and
specimen can be shifted and adjusted to the
requirements of the tissue to be sectioned.
Loosen the clamping lever (fig. 44.4) on the
right side of the knife carrier and move the
upper part of the knife carrier (fig. 44.2) on the
base (fig. 44.1).
The adjusted clearance angle can be read on
the side scale.
Then turn the clamping lever (fig. 44.4)
upwards to lock in the new clearance angle.
The clamping lever for the angle adjustment
can be pulled off after a correct angle setting to
avoid that the angle is "reset" unintendedly.
Note:
By experience, usable cuts are only
achieved at a clearance angle of 10° or
more.
If the clamping lever (fig. 44.4) is loosened, the
upper part of the knife carrier can be moved 1
cm to the left or right side.
This way, the cutting edge can optimally be
used.
Fig. 44