Instruction manual
Thermo Scientific Rotary Microtome Microm HM 340 E
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387831 - English
43
Clearance angle adjustment:
• The clearance angle between cutting edge and
specimen can be shifted and adjusted to the
requirements of the tissue to be sectioned.
• Loosen the clamping lever (fig. 43.3) on the
right side of the knife carrier and move the
upper part of the knife carrier (fig. 43.2) on the
base (fig. 43.1).
• The adjusted clearance angle can be read on
the scale.
• Then turn the clamping lever (fig. 43.3)
upwards to lock in the new clearance angle.
Note:
By experience, usable cuts are only
achieved at a clearance angle of 10° or
more.
• If the clamping lever (fig. 43.3) is loosened, the
upper part of the knife carrier can be moved 1
cm to the left or right side.
• This way, the cutting edge can optimally be
used.
• The clamping lever for the angle adjustment
can be pulled off after a correct angle setting to
avoid that the angle is "reset" unintendedly.
Moving the knife carrier on the console:
• Loosen the clamping lever (fig. 43.6) on the left
side of the knife carrier, to move the carrier
forwards or backwards on the dovetail guide.
• This allows a rough adjustment of the knife to
the specimen.
Protection against injury:
• A bracket (fig. 43.4) on the clamping plate can
be moved upwards over the blade for the
protection against injury.