3-11-2 DISPOSABLE BLADE CARRIER E
Inserting the blade:
• The disposable blade carrier E is designed to
take all commercially available high and low
profile blades.
• Insert the blade into the slot behind the
clamping plate (fig. 43.5).
• When using high profile blades, first loosen the
two screws (fig. 43.9) and remove the spacer
strip (fig. 43.3).
• Turn the clamping lever (fig. 43.7) to the front.
• Swing the bracket with scale (fig. 43.4) to the
front.
• Loosen the clamping lever (fig. 43.7) and, if
necessary, slightly press the lower part of the
clamping plate as well.
• A small gap between rail (fig. 43.3) and
clamping plate (fig. 43.5) can be seen.
• Insert the blade on the rail (fig. 43.3) and push
it from the side to the middle.
• Afterwards, return the clamping lever (fig. 43.7)
upright, thus locking the blade in position.
• The bracket (fig. 43.4) is provided with a scale.
• After loosening the clamping lever (fig. 43.7)
and after having moved the bracket (fig. 43.4)
upwards, move the blade together with
clamping plate (fig. 43.5) according to the scale
by means of the knife guard to the left or right
side.
• This way, the entire cutting length of the blade
can be used. Then press the clamping lever
(fig. 43.7) upwards.
• The levers (fig. 43.7 and 43.3) can be removed
by pulling them off towards the side.
• The lever (fig. 43.7) can also be used on the
left side. This way, the blade can be clamped
with the left hand.