Instruction manual

Thermo Scientific Rotary Microtome Microm HM 340 E
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387831 - English
41
Using high profile blades:
When using high profile blades, first remove the
spacer strip (fig. 42.3).
For this, turn the clamping lever (fig. 42.7) to
the front until the stop.
The press the stop pin into the knife carrier and
further turn the clamping lever (fig. 42.7).
Pull off the clamping lever (fig. 42.7) and
remove the clamping plate (fig. 42.3).
Insert the clamping plate and clamping lever
again.
Push the stop pin into the knife carrier and turn
the lever into an upright position.
Clearance angle adjustment:
The clearance angle between cutting edge and
specimen can be shifted and adjusted to the
respective requirements of the tissue to be
sectioned.
Loosen the clamping lever (fig. 42.8) on the
right side of the knife carrier and move the
upper part (fig. 42.9) of the knife carrier on the
base (fig. 42.1).
The adjusted clearance angle can be read on
the scale.
Then turn the clamping lever (fig. 42.8)
upwards to lock in the new clearance angle.
The clamping lever for the angle adjustment
can be pulled off after a correct angle setting to
avoid that the angle is "reset" unintendedly.
Note:
By experience, usable cuts are only
achieved at a clearance angle of 10° or
more.
Moving the knife carrier on the console:
Loosen the clamping lever (fig. 42.6) on the left
side of the knife carrier, to move the carrier
forwards and backwards on the dovetail guide.
This allows a rough adjustment of knife to
specimen.
Protection against injury:
A bracket (fig. 42.4) on the clamping plate can
be moved upwards over the blade for the
protection against injury.