3-10 SPECIMEN CLAMPING
To clamp specimens, different specimen clamping
systems are available. With the orienting adapter it
is simple to align the specimen properly in relation
to the knife.
3-10-1 UNIVERSAL CASSETTE CLAMP
The universal cassette clamp (fig. 36) represents a
quick change system.
• To insert or remove the cassette from between
the fixed (fig. 36.1) and movable (fig. 36.2)
jaws, pull the lever (fig. 36.3) to the front (in the
direction of the arrow).
Note:
To achieve optimal clamping keep the
locating surface of the cassette free of
paraffin.
3-10-2 UNIVERSAL CASSETTE CLAMP,
ADJUSTABLE
• To insert and/or remove cassettes between the
fixed jaw (fig. 37.1) and the movable jaw (fig.
37.2), move the lever (fig. 37.3) upwards.
• To adjust the size to a cassette size which is
not according to the norm, use the nut (fig.
37.4).