3-9 ADAPTERS FOR SPECIMEN
CLAMPING
3-9-1 ADAPTER, NON-ORIENTING
This adapter serves for the non-orienting fastening
of the specimen clamps directly on the cylinder of
the instrument.
3-9-2 ADAPTER, ORIENTING,
SPECIMEN ORIENTATION
Using the orienting adapter fasten the
specimen clamps on the cylinder head. This
allows the specimen to be aligned in relation to
the knife.
To bring the specimen into the desired
position, turn the clamping lever (fig. 32.1)
towards the front.
This will loosen the specimen clamp and a
rotation of 360° on the cylinder axis (Z axis) is
possible.
With the two orienting screws (fig. 32.2), the
specimen clamp can be oriented 8° in each
direction on the X axis and Y axis.
After having oriented the specimen, turn the
lever (fig. 32.1) upwards to fix the specimen
clamp in its position before starting sectioning.
When turning the orienting screws (fig. 32.2) a
slight resistance can be felt in a specific
position.
This means that the clamp is aligned in a
parallel way in relation to the cutting surface.
Fig. 32