Instruction manual

Thermo Scientific Rotary Microtome Microm HM 340 E
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387831 - English
30
3-8-3 TURNING OFF THE FUNCTION
"RETRACTION"
If desired, the function "retraction" can be turned
off.
Press the menu button (fig. 27.1).
Select submenu "Retraction" via the knob (fig.
28.1).
Press the knob to confirm the selection.
Turn the knob to select the desired function:
R:ON or R:OFF.
Press the knob to confirm the setting.
The yellow LED RETRACTION, however,
might still be on and goes off only after having
passed from the return travel again to the
cutting movement by means of the handwheel.
To turn on again the function "Retraction",
please proceed as described above.
RETRACTION R:ON
Select: Turn knob
Confirm: Press
fig. 26
Fig. 27
Fig. 28
Retraction
MEMO
MENU
1
2