3-8-3 TURNING OFF THE FUNCTION
"RETRACTION"
If desired, the function "retraction" can be turned
off.
• Press the menu button (fig. 27.1).
• Select submenu "Retraction" via the knob (fig.
28.1).
• Press the knob to confirm the selection.
• Turn the knob to select the desired function:
R:ON or R:OFF.
• Press the knob to confirm the setting.
• The yellow LED RETRACTION, however,
might still be on and goes off only after having
passed from the return travel again to the
cutting movement by means of the handwheel.
• To turn on again the function "Retraction",
please proceed as described above.
RETRACTION R:ON
Select: Turn knob
Confirm: Press