3-7 SPECIMEN FEED
3-7-1 COARSE FEED AND SPEED FOR
COARSE FEED
After changing the specimen or moving the knife or
knife carrier, it is necessary to adjust the specimen
to the knife edge again. This can easily be done by
means of the specimen coarse feed and the
defined trimming values.
For the fast forward and backward travel between
specimen and knife edge, the microtome has a
motorized coarse feed system.
The specimen holder is moved to the front as
long as the knob (fig. 18.2) for activating the
coarse feed forwards, i.e. in the direction of the
user, is turned.
The feed speed is controlled via the turn angle.
When the knob is only slightly turned forwards,
the speed is slow. Turn the knob further
towards the front, the speed will be higher.
To move the specimen holder backwards, also
turn the knob (Fig. 18.2) backwards, i.e. in the
opposite direction of the user.
In this case, the swing angle from the zero
position on is also responsible for the coarse
feed speed.
In case the knob is kept in its maximum turn
angle for more than two seconds, the
specimen movement automatically moves in its
rear end position after releasing the knob.
Then the coarse feed motor turns off after
having reached the rear end position.
When the front end position of the specimen
movement is reached, the red LED arrow on
the operating panel (fig. 19.2) lights up.
When the rear end position of the specimen
movement is reached, the red LED arrow on
the operating panel (fig. 19.1) lights up.
Press the knob (fig. 18.2) to release trim feed
with the selected value, even if the fine mode is
active.
Fig. 18
Fig. 19
Retraction
MEMO
MENU
1
2
1
2