Instruction manual

Thermo Scientific Rotary Microtome Microm HM 340 E
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387831 - English
22
3-4 SETTING SECTION THICKNESS AND
TRIMMING THICKNESS
The required section and trimming thicknesses
are set by means of the control knob (fig. 9.1),
which is placed on the side part of the
operating panel.
To choose between section thickness and
trimming thickness, press the control knob (fig.
9.1). The selected thickness range is then
marked with a rim when in the operating mode
NORMAL.
Note:
Whenever the instrument is turned on,
"FINE" sectioning is automatically
selected and shown accordingly (fig.
10.1).
FINE = pre-selected section thickness
TRIM = pre-selected trimming thickness
The graduation of the section thicknesses is
divided into six ranges:
Range: Graduation:
up to 2 µm 0,5 µm
from 2 µm to 10 µm 1 µm
from 10 µm to 20 µm 2 µm
from 20 µm to 30 µm 5 µm
from 30 µm to 40 µm 10 µm
from 40 µm to 100 µm 20 µm
The graduation of the trimming thicknesses is
divided into four ranges:
Range: Graduation:
up to 10 µm 5 µm
from 10 µm to 100 µm 10 µm
from 100 µm to 200 µm 20 µm
from 200 µm to 500 µm 50 µm
Fig. 9
Fig. 10
Retraction
MEMO
MENU
1
2
FINE= 9
TRIM= 30
SUM = 0
R : ON
0
1