3 OPERATING INSTRUCTIONS
3-1 SETTING UP THE MICROTOME
Cut through the bands around the carton.
Open the carton.
Remove the accessories.
On the lower rear side of the instrument, there
is a recessed grip to lift or carry the
microtome.
Lift the microtome out of the carton.
Note:
Do not use the handwheel handles to
lift or carry the instrument.
Choose the site of installation in a way that the
switch for separating the instrument from the
voltage can be reached at any time.
Place the microtome on a stable and vibration
free table, as sectioning can be influenced by
nearby instruments which generate vibrations.
Handwheel and coarse feed wheel must be
free and accessible in a comfortable way.
Remove the separately packed section waste
tray and install it at the base plate from the
front side.
In the rear part of the base plate, there are
sliding feed for an easy moving of the
microtome.
Slightly lift the base at the front end only and
slide the microtome.
Note:
Remove the section waste tray to
move or carry the instrument. The
section waste tray can be pulled out of
its proper position.
Before starting section, instrument, knife
carrier and section waste tray should be
treated with a commercially available paraffin
repellent.
Note:
This medium considerable reduces the
adhesive force of paraffin sections on
the individual parts (see part 2-2-2-5,
optional accessories).