Thermo Scientific Rotary Microtome Microm HM 340 E
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387831 - English
16
2-3 TECHNICAL DATA HM 340 E
Microtome: Section thickness range ......................................... 0,5 - 100 µm
Resolution .............................................. 0,5 µm from 0,5 - 2 µm
................................................................... 1 µm from 2 - 10 µm
................................................................. 2 µm from 10 - 20 µm
................................................................. 5 µm from 20 - 30 µm
............................................................... 10 µm from 30 - 40 µm
.............................................................. 20 µm from 40 - 100 µm
Trimming thickness range...........................................5 - 500 µm
Resolution ...................................................5 µm from 5 - 10 µm
.............................................................. 10 µm from 10 - 100 µm
............................................................ 20 µm from 100 - 200 µm
............................................................ 50 µm from 200 - 500 µm
Specimen retraction during return travel ........................... 40 µm
Horizontal feed range ...................................................... 28 mm
Vertical specimen stroke ................................................. 64 mm
Section counter: ......................................................................... 5-digit, with reset
Section thickness sum: ......................................................................... 5-digit, with reset
Remaining travel
to front end position: ..........................................................................................5-digit
Specimen size: when using a standard specimen clamp: .................. 55 x 50 mm
Specimen orientation: x - and y - axes........................................................ universal 8°
rotation:...................................................................... up to 360°
Cutting drive: ................................................. manual by means of handwheel
Coarse feed: ......................................... motorized, graduated and continuous
Speed for
coarse feed: ................................................................ 400, 800 or 1200 µm/s
Storage
temperature range: ...................................................................... -20°C up to +50°C
Operating conditions: .......................+5°C up to +40°C (at a max. rel. humidity of 60%)
............................................................altitude up to 2000 M.S.L.
.......................................................................for indoor use only
Floor loading requirements:.........................................................................................103 kg
Power requirements: ........................................220...230 V 200 mA +/-10% 50...60 Hz
................................................ 240 V 200 mA +/-10% 50...60 Hz
................................................ 100 V 400 mA +/-10% 50...60 Hz
................................................ 115 V 400 mA +/-10% 50...60 Hz
Pollution degree: ................................................................................................. 2
Overvoltage category: ................................................................................................. II
Sound pressure: ......................................................................................39 dB(A)
Dimensions: ........................... 410 mm x 520 mm x 280 mm (wide/deep/high)
Weight: ...........................................................................................28 kg