After the transmission of your record(s), the Real-Time ECG View is displayed.
Direct Connection
For a direct connection hookup, set the Default XMT media to RS232 from Configuration page 6 (see section 3 of
this manual). Connect the ELI 250 to another Mortara Instrument electrocardiograph, to the E-Scribe, or to ELI
LINK with a direct connect serial cable.
In the configuration setting, select matching baud rates for both units. Use 38400 bps for a direct connection to E-
Scribe.
Modem Transmission
For a modem transmission, set the Default XMT media to Modem from Configuration page 6 (see section 3 of this
manual). Connect the ELI 250 to a standard telephone jack with the provided phone line cable. Plug the cable into
the telephone jack located on the back of the cardiograph and the other end into a telephone wall jack. Confirm
telephone number in the configuration settings (see Section 3 of this manual).
Modem Initialization
The modem initialization string is country specific. At the time of production, the modem initialization string is
configured for the country of purchase. However, if the unit is relocated to a different country, the modem
initialization string will need to be modified.
From the Real Time ECG View, select F6 (More) to display the Application Menu.
Real Time ECG View
Application Menu