• As defined by IEC 60601-1 and IEC 60601-2-25, the device is classified as follows:
• Class I equipment or internally powered
• Type CF applied parts
• Ordinary equipment
• Not suitable for use in the presence of flammable anesthetics
• Continuous operation
NOTE: From a safety perspective, per IEC 60601-1 and derivative standards / norms, this unit is declared to be
“Class I” and uses a three-prong inlet to ensure an earth connection is made along with mains. The ground terminal
on the mains inlet is the only protective earth point in the unit. Exposed metal accessible during normal operation is
double insulated from mains. Internal connections to earth ground are functional earth.
• The ELI 250/210 will automatically turn off (blank screen) if the batteries have been severely discharged.
• After operating the ELI 250/210 using battery power, always reconnect the power cord. This ensures that the batteries will
be automatically recharged for the next time you use the ELI 250/210. A light will illuminate, next to the on/off switch,
indicating that the unit is charging.
• The ELI 250/210 is a UL Classified Device:
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH
UL2601-1, IEC60601-1, CAN/CSA CC22.2 No. 601.1, AND
IEC60601-2-25