Service manual

ELI 250/210__________________________________________________________________________
3-24
Configuration Page 2
Date Format
This option defines the format for entering and displaying the patient’s date of birth in either the U.S. format or the
European format.
For example: July 16
th
, 2003 is displayed as follows:
US Date Format: MM/DD/YYYY (07/16/2003)
European Date Format: DD.MM.YYYY (16.07.2003)
NOTE: the date format option does not modify the acquisition date printed on each ECG. Using the
example above, the acquisition date would print as 16-Jul-2003.
Plot Format
This option defines the default for one of five available plot formats in either Standard or Cabrera presentation.
Please note that regardless of the plot format selected, 10 seconds of 12-leads are always stored and printed.
The ECG plot options are:
Format
Option
ECG Data
3+1
2.5 seconds of 12 leads in a three-channel format
and the fourth channel is a 10 second rhythm strip
of lead II (user defined).
6
5 seconds of 12-leads in a six-channel format
3+3
2.5 seconds of 12 leads in a three-channel format
plus 10 second rhythm strip of leads II, V1 and V5
(user defined) in a three-channel format.