This option defines the number of ECGs retrieved from an E-Scribe data management system. The ECGs are
retrieved by ID number. A zero (0) setting retrieves the most current ECG for that ID number. Settings from one
(1) to nine (9) retrieve the most current ECG plus “X” number of ECGs identified by the entered value.
EXAMPLE: If you enter the number 5, you will retrieve the most current ECG plus the five preceding ECGs for
that ID number. ECGs retrieved from the E-Scribe are only printed at the ELI250 and not saved.
Delete Rule
This option defines the ECG auto delete status
in the patient directory. ECGs that are marked for deletion will be
automatically removed or erased based on their record size to make room necessary for the new ECG record. ECGs
are only erased from the directory when they are marked for deletion and if the directory becomes full. The delete
rule selections are:
Post Plot = ECG is automatically marked deleted after printing.
Post Transmit = ECG is automatically marked deleted after transmission.
Post Plot/Transmit = ECG is automatically marked deleted after transmission and printing.
Storage Sensitivity
The ECG storage sensitivity setting dictates the resolution of the stored record. The sensitivity settings are either
Normal or High. If the configuration value is set to High, the stored ECG will have a high resolution. As a result,
the records size will be large and will reduce the storage capacity in the ECG directory.
Auto Save ECG
This option defines whether or not the ELI 250 is to automatically save newly acquired ECGs in the directory, once
they are acquired and printed. If the auto save configuration option is set to NO, and the record is printed, the ELI
250 will prompt you to “Save ECG?” SAVE (F1) will store the ECG in the directory.
NOTE: manually save ECGs by selecting More (F5) from the Acquired ECG view and Save (F5).
Auto Print ECG
This option defines whether or not the ELI 250 will automatically print the ECG after acquisition. If the selected
configuration option is set to NO, a manual printout is possible after acquisition.
Baud Rate
The user-selectable baud rate determines the serial port’s data transmission rate in bits per second (bps). Set the
baud rate to: 9600, 19200, 38400, 57600, or 115200 bps for direct data transmission between the ELI 250 and
another Mortara cardiograph, and 38400 bps for a direct connection to the E-SCRIBE data management system.
Use A4 Paper
The ELI 250 accommodates use of Z-fold thermal paper in either the letter size (8.5 x 11 inches; 216 x 279 mm) or
the “A4” size (8.27 x 11.69 inches; 210 x 297 mm). The provided paper tray spacer is required for use with A4 size
paper. Enable this configuration option for A4 printing. The paper tray spacer should not be inserted if the letter
size paper is used.
Caps Lock
All character entry is translated to upper case.