“Waiting for Response”, “Connected”, and “Custom ID downloaded” will be displayed before returning to the Real-
Time ECG View, which indicates the Custom ID download is complete. The new Custom ID becomes the
customized header format for all future ECGs until you select a different ID format in the configuration settings.
You may alter the ID format configuration to Short, Standard, Long or Custom based on your patient demographic
entry needs. The Custom ID is only deleted upon downloading a new Custom ID or on the rare occasion of
downloading software – it will not be lost due to power loss or switching to a different ID format.
TIP: Upon Custom ID download, the ID format will assume the group name as designed in the E-Scribe.
NOTE: A Custom ID must be downloaded before downloading the Request Codes (please reference the E-Scribe
Operator Manual).
NOTE: The site number must be configured in the cardiograph and recognized as an established, valid site
number at the E-Scribe before downloading the custom ID.
TIP: Confirm the baud rate in the configuration settings before downloading the custom ID from the E-Scribe.