Instruction manual
81 
Possible cause: 
f)  Changeover times and voltage levels are outside recommended 
  tolerances for emergency lighting systems with changeover 
  between AC and DC 
  Remedy: 
  Measure the DC supply voltge and check the switchover 
  properties, or consult the equipment manufacturers. 
Possible cause: 
g)  Non-sinusoidal mains voltage or DC voltage with high residual 
  ripple (e.g. operation with fully electronic leading edge phase 
  control dimmer switch or artificial mains network/emergency 
 generating set) 
 Remedy: 
  Check whether the mains voltage is in the required range for the 
  ECG and the wave form or residual ripple in DC operation is within 
  the specified limit values. Dimming is not recommended unless the 
  ECG is expressly approved for dimming, in which case the 
  prescribed controls (special accessories) must be used. 
Possible cause: 
h)  At least one lamp has reached the end of its life owing to a increase 
  in lamp voltage 
  Remedy: 
  Replace the lamp (for two-lamp ECGs we recommend replacing 
  both lamps at the same time to keep maintenance time and costs 
  to a minimum) 
8.5.3  Lamp goes out during 
operation 
Problem: 
The lamp goes out during operation (at least one lamp in the case of two-
lamp units) 
Possible cause: 
a)  The reversible protection device in the ECG has responded during 
  operation. The reason may be an intensive transient mains fault 
  (mains voltage falls below the minimum voltage specified on the 
  ECG for longer than permissible). Pulses of exceptional energy 
  (transients) are superimposed on the mains voltage. The value of 
  the mains voltage exceeds the maximum recommended value (e.g. 
  because of a fault in the supply unit). Slow increase in mains 
  voltage following neutral conductor break (unbalanced load, 
  depending, among other things, on the mains load). 
  Remedy: 
  Disconnect the ECG or luminaire from the mains then check the 
  supply voltage. If such problems occur sporadically, we 
  recommend recording the mains voltage and/or using an 
  oscilloscope or memory voltmeter. The electricity supplier may 
  have to be consulted. Make sure that the neutral conductor is 
  connected properly to this luminaire and makes good contact. 










