Instruction manual
78 
If a QUICKTRONIC
®
-ECG for T5/∅ 16m m-fluorescent lamps is operated 
over long periods on a supply voltage greater than 280 V it may fail as a 
result of overheating (with the exception of QUICKTRONIC
®
INTELLIGENT). 
8.3 Equipment Behaviour 
on Under Voltage 
Important: 
Significant under voltage can cause ECG failure for all ECG in constant 
wattage circuitry. Due to the characteristics of arrangement the line current 
increases for decreasing voltage. Below the specified voltage range this 
causes an overload of the filter chokes. First the colour of the copper wire 
and board below are changing. For a longer overload both windings of one 
or both filter chokes fail due to melting or swelling in the windings. 
Compact and straight fluorescent lamps go out below the specified value of 
the ECG or even don’t ignite if voltage is too low. The ECG will not be 
affected while the lamp is still burning. If the lamp is disabled at the end of 
its life (i.e. emitter paste is wasted), the safety shut-down of the ECG 
cannot work below the specified voltage range and a damage of the ECG is 
possible. 
8.4 Application faults   
8.4.1  Wiring faults on the 
lamp side 
All single-lamp QUICKTRONIC
®
 ECGs and units with 6-pole output 
terminals: 
If there is a wiring fault the lamp will not operate or will cold start. In 
particularly serious cases there is a risk that the ECG will be destroyed. 
8.4.2  Short-to-ground at the 
output of 
QUICKTRONIC
®
 ECG 
If there is a short-circuit-to earth at one or more of the connecting cables 
between the ECG and the lamp, the ECG will fail. 
Damage to the ECG: 
•  Melting and swelling of only one winding on one or both input filter 
chokes 
•  Breakdown of one or two rectifier diodes, irrespective of whether 
they are discrete or integrated components. 
There are, however, other causes of these major faults. Generally, as a 
consequence, consequential damage occurs at other components and this 
damage need not be unique in character. 
8.4.3  Effects of moisture 
Rust at the edges of metal housings may indicate the effects of moisture. If 
the printed circuit board and/or various components are tarnished, it can be 
assumed that there has been a serious moisture ingress. There will be 
short-circuits between neighbouring component connections or solder 
points with high potential difference, resulting in failure of the ECG. 
Symptoms of serious faults are indicated by a “tidermark” inside the 
housing. 
8.4.4 Installing luminaires 
in draughty locations 
Depending on the location of the fluorescent lamp in the draught, the lamp 
will cool down in certain areas of the bulb. This leads to local depletion of 
mercury and to a darkening of the lamp in these areas because there is no 










