Instruction manual

45 
significant change in temperature for some time). The supply voltage 
should be held constant at least throughout the entire measuring cycle at 
the rated voltage of the lamp. 
The 
following procedure is recommended for the thermal analysis of the 
luminaire, taking into account the design requirements specified in EN 
60598-1: 
1.  Thermal situation in the luminaire without contol gear heat. 
Luminaire in measurement setup according to EN 60598-1 in 
standard mounting position, equipped with ECG and lamp and fitted 
with thermocouples. The lamp is supplied from external control 
gear, and not from the built-in ballast. In this way the temperature 
rise in the entire set-up resulting only from the lamp can be 
measured and the thermal “link” to the environment can be 
optimized. 
2. 
Thermal situation in the luminaire with contol gear heat. 
Arrangement as described in 1., but the lamp is supplied from 
internal control gear. By comparison with the measured values 
obtained already, the additional heat generated by the ECG can 
now be assessed. 
3.10  Luminaire Wiring Test 
for Two-lamp 
Luminaires 
3.10.1  Testing with a Test 
Adapter and Dummy 
Lamps 
A more precise wiring test for two-lamp luminaries than the one described 
in Section 3.10.1 can be performed with a test adapter (own design with 
the resistors shown in the diagram) and a sample tube (dummy lamp with 
sockets measuring the resistance). This test can be used for two-lamp 
luminaires equipped with the before mentioned ECGs. 
¾Insulation test: 500 VDC acc. to EN 60598 app. 3...5 sec.; R = 2MΩ
¾High voltage test (100% for each luminaire)
N
L
1
2
3
4
QUICKTRONIC
®
lamp terminal
output
socket 1
output
socket 2
Earth connection via ECG- housing 
¾Insulation test: 500 VDC acc. to EN 60598 app. 3...5 sec.; R = 2MΩ
¾High voltage test (100% for each luminaire)
N
L
1
2
3
4
QUICKTRONIC
®
lamp terminal
output
socket 1
output
socket 2
Earth connection via ECG- housing 










