Instruction manual
22 
system. The cut-off of the permanent filament heating after lamp ignition is 
an advantage. 
Further the values of the horseshoe curves also indicate the arrangement 
of the lamps within the fixture. To avoid thermal influences of the lamps 
minimum distances in between have to be kept. The lamps have to be 
placed that the stamp of all lamps is on the same side. For vertical 
arrangement the stamp of the lamp always should be placed down. For 
circular lamps FC the socket has to be placed down. 
The measurement of the cold spot temperature is especially important for 
the luminaire manufacturer. This temperature offers opportunities to 
optimize the luminaire efficacy. 
2.15  End of Life 
(EoL acc. to T.2) 
Fluorescent lamps use up their emitter during operation. The complete loss 
of emitter on an electrode results in a voltage increase in the vicinity of the 
electrode filament. As most Electronic control gear have no problem 
providing this high asymmetric voltage and with it the required additional 
power the temperature around the electrodes rises significantly. 
At the moment the international ECG safety standard IEC 61347-2-3 is 
under revision. In the future, all ECG must provide a “end-of-life” safety 
shutdown which is continuously controlled in order to avoid overheating of 
lamp sockets. 
The actual status of the standard considers three different test circuits for 
Electronic control gear. If an ECG complies with one of the three tests, it 
offers protection against the “end-of-life” behavior of fluorescent lamps. 
•  Asymmetric pulse test    (Test 1) 
•  Asymmetric power test    (Test 2) 
•  Open filament test      (Test 3) 
The asymmetric power test (Test 2) is directly simulating the additional load 
which results from the increased asymmetric voltage in the case of “end-of-
life”. In order to pass the test the additional load may nod exceed a specific 
value depending on the lamp type. Most ECG experts see test 2 
(asymmetric power test) as the safest “end-of-life” control, because the 
direct measurement of the asymmetric additional load mirrors the real lamp 
behavior at its end of life. OSRAM does not compromise the safety of 
Electronic control gear and has, for quite some time, specified Test no. 2 as 
standard test. 
2.16 U-OUT 
U-OUT a binding ECG label according to safety standard EN 61347-2-
3. The former standard EN 60928, still valid until 2006, allows labeling 
of U-OUT either in the product catalogue or on the ECG housing. U-
OUT specifies the largest effective working voltage between 
-  Lamp terminals 
-  Each lamp terminal and earth connection, if applicable 
This information is important for all components electrically connected on 
the lamp side of the ECG. 
All components such as lamp cables, sockets (EN 60061-2), isolating 










