User guide
CS 9300 System User Guide (SM748)_Ed 01 8–1
Chapter 8
TROUBLESHOOTING
Quick Troubleshooting
Occasionally, malfunctions can occur during use in the event of an incorrect action. An information (I) error code
is displayed on the Display Screen of the unit Control Panel and the message is displayed on the popup on the
Acquisition interface System Status Screen. In some cases, an audible warning is also issued.
The following table lists the information messages, their description and the action to take:
IMPORTANT
If and “E” message is displayed, the malfunction persists or
more serious conditions occur, contact a qualified technician.
When you call the qualified technician have the following
information ready:
Model Number: CS 9300
Error Code Number: E xxx
Message displayed on the popup on the Acquisition
interface.
Table 8–1 Error Message
Information Error Code Error Message Description Action
E 1 Handswitch was released
before the end of
exposure
The user has released the
exposure button prematurely.
Relaunch the acquisition and hold the
exposure button until the end of the
acquisition cycle.
Table 8–2 Quick Troubleshooting
Information Error Code Information Message Description Action
I 1 X-Ray tube cooling Cooling in progress. Wait until the Generator Cooling Indicator
on the Acquisition interface reaches zero.
I 2 Thermal security Cooling in progress. Wait until the Generator Cooling Indicator
on the Acquisition interface reaches zero.
I 3 Release handswitch The acquisition has ended. Release the exposure button of the x-ray
remote control.