User guide
Controlling the Image Quality
DRAFT CS 9300 System User Guide (SM748)_Ed 01 7–3
3. In the Too l s pane, double-click Quality Assurance Control. The Image Quality
Control window is displayed.
4. In the Image Quality Control window, to perform the test, you can either:
Select the desired device in the List of detected acquisition devices field.
Or
Browse and import a file in the Import field.
Click Next and follow the instructions.
5. Select the desired test and follow the on-screen instructions.
You have two test options:
Constancy test: This test must be executed regularly by the user. The result of
the tests must be compared to the Acceptance test. Select the desired test and
follow the on-screen instructions.
Or
Acceptance test: This test is ONLY executed by the technician. It is a complete
image quality control test (make sure you have the required test tools). The result
of this test is used as reference for the Constancy test by the user.