User guide
CS 9300 System User Guide (SM748)_Ed 01 7–1
Chapter 7
MAINTENANCE
Cleaning Schedule
This section describes the cleaning schedule that you need to perform regularly for your
CS 9300 unit and the accessories.
Daily
Carry out the following maintenance tasks:
WARNING
Switch off the unit, then, clean all accessible parts of the
machine with a detergent product or a detergent-disinfectant.
Avoid using liquids inside the unit. Follow the detergent
product or the detergent-disinfectant manufacturer
recommendations for safety precautions.
CAUTION
We recommend the use of barrier protected accessories,
available from distributors, between each patient use. If no
barrier protection is used, follow the daily maintenance tasks
described below. Use the disinfectant product according to
the recommendations of the manufacturer for safety
precautions.
Table 7–1 Daily Maintenance Tasks
Accessories Maintenance Tasks
3D and Panoramic bite block Sterilize with cold sterilization or autoclave before the next patient is
x-rayed.
Edentulous bite block Sterilize with cold sterilization or autoclave before the next patient is
x-rayed.
Temple support Disinfect the temple support and chin rest with either an intermediate
or low level disinfectant after each use.
Chin rest (panoramic, sinus and
TMJ)
3D head rest
All components that come into
contact with the patient and the
operator
Clean all components with either an intermediate disinfection or
low-level disinfection between each patient use.
Outer covers of the unit Wipe the unit with a dry cloth at the end of each day's operation.
WARNING
Do not use detergents or solvents to clean the outer covers of
the unit.