User guide

Acquiring a TMJ x4 Image
CS 9300 System User Guide (SM748)_Ed 01 5–7
Acquiring a TMJ x4 Image
Before acquiring a TMJ x4 image, check that you have:
Reset the unit rotative arm to the start position for patient to enter the unit.
Selected the patient record.
Accessed the Imaging window.
Accessed the Panoramic Acquisition interface.
To acquire a TMJ x4 image, follow these steps:
1. From the Panoramic Acquisition interface, in the Program pane, select the TMJ and
click .
2. Remove , place on the chin rest support and cover it with a hygienic
barrier.
3. Press and hold to raise the chin rest to the TMJ x4 positioning indicator
.
4. Correctly position the patient using the 2 positioning laser beams.
5. Ask the patient to remain still and close the eyes. Acquire an image with the mouth
closed.
6. To acquire an image with open mouth, on the Control Panel, click to reset the
unit rotative arm.
7. Ask the patient to stay in the same position and open the mouth. Acquire an image
with the mouth open.
8. Do the following when the acquisition is finished:
Open the temple supports and release the patient.
Remove the hygiene barrier of the bite block.
Reset the unit rotative arm for the next acquisition.