User guide

Acquiring a Panoramic, TMJ x2 or Sinus Image
5–6 ACQUIRING PANORAMIC IMAGES
Launching the X-ray
To launch the x-ray, follow these steps:
1. Leave the x-ray room and close the door. You must keep visual contact with the
patient during acquisition.
2. Launch the x-ray with the remote control. Press and hold the exposure button until the
end of the acquisition is indicated by the on-screen “Release Switch” message. The
turns yellow and a warning sound is heard, indicating x-ray emission. When the
acquisition ends, the Panoramic Acquisition interface disappears and the acquired
image is automatically transferred to the Imaging window.
3. Check the image quality.
4. Do the following when the acquisition is finished:
Open the temple supports and release the patient.
Remove the hygiene barrier of the bite block.
Reset the unit rotative arm for the next acquisition.
IMPORTANT
To stop the acquisition, if any problem, release the exposure
button of the remote control or press the red emergency stop
button.