CS 9300 Imaging System User Guide (SM748)_Ed 01 4–1
Chapter 4
GETTING STARTED
Switching on the Unit
Before switching on the unit, check that:
The installation of the unit is complete.
The computer is switched on.
IMPORTANT
You must switch on the computer and wait for it to be ready for the
connection before switching on the unit.
To switch on the unit, follow these steps:
1 On the unit column, press the ON button.
2 Switch on the unit and wait for a minute for the connection between the unit and the
computer to be established. In the status bar, must become .
3 If you start the imaging software before the connection is established an error
message is displayed. Click OK, close the imaging software and wait for the
connection to be established.
4 You can now proceed to start the imaging software.
IMPORTANT
To increase the operating life of the x-ray tube, if the unit has not been used
for a month, you must follow the following procedures before use.
1 In the Panoramic Acquisition interface, select the Parameter pane.
2 Select the following series of parameter settings:
70 kV - 6.3 mA
80 kV - 10 mA
85 kV - 10 mA