User guide
Acquisition Interface
CS 9300 System User Guide (SM748)_Ed 01 3–7
“Premium” 3D Acquisition Overview
The “Premium” 3D Acquisition interface is the optimum 3D interface for the CS 9300
that provides you with the full range of imaging acquisition functions.
Figure 3–5 “Premium” 3D Acquisition interface
1 Information button:
About: Identifies the Software and the Firmware versions.
Reset of the Values: Resets to the manufacturing parameter settings.
Memorize settings: Memorizes the user preference settings for each patient type (kV,
mA and voxel size).
Save current mode as default: Sets the current selected program to be available by
default each time you access the Acquisition interface.
Activate the Phantom mode: Sets the unit parameters to perform an image quality
control test.
2 Preview Screen:
Before exposure: Displays an illustration of the selected FoV inside the anatomy.
During exposure: Displays the acquired image in real time.
3 Selected Parameter Display: Displays the current acquisition parameter settings.
4 System Status Screen: Displays various alert or warning messages originating from the unit.
5 Generator Cooling indicator: Indicates the automatic cooling time (mm:ss) required for the
generator to reach 0 for a new acquisition.