9 Troubleshooting
9.1 Table of Error Messages
Error Message
Description
Resolution
Start Problem
Instrument busy Trying to enter a command
while the instrument is still
processing another command
Wait until the task is finished then re-
enter the command
Injector not
initialized
Injector is needed for the
protocol but was not primed
before starting the run
Touch the
SETUP
button and follow
the wizard to PRIME the injector
No plate in
instrument
The instrument does not
detect a microplate in the
Microplate Sample Tray
Make sure there is a microplate seated
properly in the Microplate Sample Tray
Protocol Selection
No wells selected
At least one well on the
microplate needs to be
selected (green) to be read
Touch the PLATE button found on the
READ screen and verify at least one
well is selected
Instrument Busy
Stop the run in
progress before
creating a new
protocol
When trying to use any of the
PROTOCOL wizards during a
run
Wait until the current run is over, then
create or modify a protocol
Instrument Problem
Door opened
Instrument door was opened
during a run causing the
reading to abort
Restart the run
Something prevented the
instrument door from fully
closing as the Microplate
Sample Tray was retracting
Check the instrument door for any
obstructions
No tray in
instrument
Prime and flushing injectors
require use of the Waste
Collection Tray
Touch the DOOR button to open door
and place the Waste Collection Tray
on the Microplate Sample Tray
Plate location
misaligned
The sensor did not detect the
sample plate
Push the Microplate Sample Tray all
the way in, close the instrument door,
and restart run
Injector 1 not
installed
The protocol requires addition
of reagent using Injector 1 but
it is not detected
Verify Injector 1 is installed correctly by
checking all connections and looking
for an INJECTOR icon on the
INSTRUMENT STATUS bar