System information

5 Injector System
www.turnerbiosystems.com 39
5.8 Removing or Replacing Inlet and Outlet Plastic Tubing
Replacement tubing assembly is available from Turner BioSystems. See Appendix C for ordering
information.
1. Disconnect the inlet or outlet tubing from the injector pump by twisting the gray fitting
counterclockwise.
2. Discard the used injector tubing assembly.
3. Twist the fitting of the replacement tubing into the correct port of the injector pump.
5.9 Removing or Replacing Stainless Steel Tubing
Replacement stainless steel tubing is available from Turner BioSystems as part of the inlet tubing
assembly. See Appendix C for ordering Information.
1. Hold the tan colored adapter just above the stainless steel tube.
2. Turn the adapter counterclockwise to disconnect it from the gray fitting.
3. Twist and separate the two tan-colored adapters.
4. Pull the stainless steel tube out of the adapter and discard.
5. Insert a new stainless steel tube into the adapter. Allow the tip of the tube to hang over the
edge approximately 3 - 4mm.
6. Twist the second part of the adapter on until hand tight.
7. Attach the adapter to the gray fitting by turning it clockwise until it is hand tight.