System information
3 Fluorescence
www.turnerbiosystems.com 24
3.3 Starting a Run
Easy-to-use protocol wizards can be accessed from the HOME screen to set up the run
parameters. Follow the steps below to set up a Fluorescence run.
1. From the HOME screen, touch the INSTRUMENT CONTROL button, and then the READ
button.
2. Toggle the Detection Mode to Fluorescence. See Figure 3-1(A).
3. Touch the white text button to select a protocol or modify the displayed parameters. See
Figure 3-1(B).
4. Touch the PLATE button and select the wells to be read. Green is selected and gray is
deselected. See Figure 3-1(C).
5. Touch the DOOR button to automatically open the instrument door. See Figure 3-1(D).
6. Place the sample plate on the Microplate Sample Tray. The A1 well must be at the top right
corner.
7. Touch the DOOR button again to automatically retract the sample plate back into the
instrument.
8. Touch the START button. See Figure 3-1(E).
9. To name the results file, touch the EDIT button to activate a keypad. Enter the results file
name and touch the OK button when editing is complete.
10. Touch the OK button on the dialog screen to save. The Modulus™ Microplate Reader will
initialize and begin run.
11. To stop a run at any time, touch the STOP button.
12. Remove the sample plate after the run is complete.
REMINDER: To prevent dehydration and spills,
remove the sample plate from the
instrument when a reading is complete.