Service manual

Revision D
3-6
Maintenance
: Domestic Electrical Safety Tests
Series 2000 Treadmill
409110-004
Leakage Tests
The leakage tests are safety tests to ensure that the equipment poses no
electrical health hazards. Use the table below to determine which tests
apply to the unit under test and the maximum allowable leakage
currents. For international leakage limits, refer to the internal standards
agencies of that particular country.
If the unit under test fails the leakage tests, do not allow the customer to
use the equipment. Call Tech Support for assistance. (See the “How to
Reach Us” page in the front of the manual.)
Marquette Medical Systems recommends that you perform these tests:
Before applying power for the first time
Every 6 months as part of routine maintenance
Whenever internal assemblies are serviced
NOTE
The accuracy of the leakage tests depends
on a properly-wired wall outlet. Do not
proceed until you verify the integrity of the
power source.
WARNING
M15287-76A
Total system leakage
current must not
exceed 300
microamperes.
Table 3-1. Leakage Tests and Maximum Allowable Leakage Currents
Test Applies To Maximum Current (µA)
1 Ground-wire-leakage-to-ground MAX-1 stress system 300
2 Chassis-leakage-to-ground MAX-1 stress system 300
3 Patient-cable-leakage-to-ground Acquisition module 10
4 Patient-cable-leakage-into-patient-leads-from-120 V ac Acquisition module 20