Service manual

Revision D
B-4
Appendix B: Technical Description
: Technical Specifications
Series 2000 Treadmill
409110-004
Table B-4. Safety
Item Description
Certification UL classified
CSA certified
Type of Protection Against Electrical
Shock
Class 1
Degree of Protection Against Ingress
of Liquids
Ordinary
Handling of Disposable Supplies and
Other Consumables
Use only parts and accessories manufactured or recommended by Marquette.
Follow manufacturer’s instructions for use for disposable/consumable product.
Follow local environmental guidelines concerning the disposal of hazardous materials
(e.g. lead acid batteries).
Patient Mode of Operation Continuous
Patient Leakage Current Not applicable
Degree of Protection Against
Electrical Shock
Not applicable
Maintenance Frequency
Test the stop switch assembly monthly.
Recommended user daily visual inspection and cleaning.
Recommended six-month routine maintenance checks and test procedures performed
by qualified technical personnel.
Repair Guidelines Calibration instructions, equipment descriptions, and all other service information to
repair those parts of the equipment designated as field repairable by qualified
technical personnel is available in the service manual.
Upon request, Marquette will provide circuit diagrams and component parts lists for
printed circuit boards deemed repairable by qualified technical personnel.
Table B-5. Environmental
Item Description
Operating Conditions
Ambient temperature 10°C to 40°C (50°F to 104°F)
Relative humidity 20% to 90%
Atmosphere pressure 525 mmHg to 795 mmHg (20.7 to 31.3 in Hg)
Storage Conditions
Temperature range –40°C to 70°C (–40°F to 160°F)
Relative humidity 15% to 95%
Atmosphere pressure 500 hPa to 1060 hPa (375 mmHg to 795 mmHg)