User`s manual

4-6
Positional Deviation Measurement
4
Register a reference image
After setting the reference image area, store an
image in the controller as a reference image.
Select " REFERENCE IMG" using the up/
down arrow keys and select "REG" from the
popup menu.
SEARCH ARE
Specify the search area (inside the dotted lines)
on the " SEARCH ARE (MDL0)" line, using the
same procedures used for setting the reference
image area.
If you are selecting a 2-point search, also set the
points the same way as for MDL1.
Note: When setting the horizontal and vertical lines, specify the line length as follows:
The reference image must shorter than the search area lines.
RECTANGLE RESET RETURN F C1 BRT
MDL
0
1
UP.L (224.208)
2
LO.R (287.271)
3
MOVE
4
REFERENCE IMG
CONTRAST SRCH
REG
DISP
X:224 Y:208
SET=MOVE ESC=BACK SEL=CHNG IMG TRG=FUNC
Object to measure
+
REG COND
   
SCREEN COND
SAVE
DETAIL
1
REGISTER NO.
0(07)
2
MEAS SHAPE (MDL0) RECTANGLE
3
REFIMG ARE (MDL0) SET224.208)(287.271
4
SEARCH ARE (MDL0) SET216.200)(295.279
SEL=VALUE INPUT ESC=BACK SEL=CHNG IMG TRG=FUNC
F C1 BRT
Search area
Model 0
Reference image
(Model 0)
Search area
(Model 0)
Model 1
Reference image (model 1)
Search area (model 1)
An example of storing
- This is an example of a 2-point search when the detection shape is a rectangle.
HCRSTSARTNOC
)hcraestsartnoc(
aeraehtrofhcraesyllacitamotua,hcraesyargaotgnignahcretfA
.egamiderutpacehtni)tsartnocmumixam(serutaeftsomehthtiw
)retsiger(GER .egamiecnereferasaegamideyalpsidyltnerrucehterotS
)yalpsid(PSID
ehtgnisusegamiecnereferderetsigerehtfoenoynatceleS
eblliwegamiecnereferdetceleseht,syekworrathgir/tfel/nwod/pu
.neercsehtfothgirrewolehtnideyalpsid