User`s manual

Table Of Contents
7-29
Simplified Menu Operation
7
[3] Initialization
If you initialize the IV-S30, the specified measurement conditions, evaluation conditions, operation
conditions, and system conditions will be reset to their initial condition.
The system time data will not be initialized.
The IV-S30 has two initialization modes: "Initialize all" and "Initialize RAM."
ËË
ËË
Ë Operating procedure
1. Move the cursor to the "4 INITIALIZATION" item using the up and down keys, and press the
SET key.
2. Select either "ALL-INIT" or "INIT-RAM" using the left and right keys.
3. Press the SET key.
- The following message will be displayed.
INITIALIZE DATA? (Do you want to initialize the data?) (YES=[SET]/NO=[ESC])
4. Press the SET key again. (Press the ESC key to abort the initialization.)
- Initialization will start, and the progress of the initialization will be displayed on the bottom
of the screen.
When initialization is complete, the message "INITIALIZING" will change to "COMPLETE INIT."
noitazilaitinI sliateD
TINI-LLA
)llaezilaitinI(
sgnittesehtsnruteR.MARdnayromemhsalfehthtobniatadehtsezilaitinI
sawenihcamehtnehwsaemaseht(sgnittestluafedyrotcafehtot
noitazilaitinisihtretfaniagaNOdenrutsirewopehtnehW.)dereviled
.dezilaitinieblliwatadeht,detcelessidohtem
MAR-TINI
)MARezilaitinI(
.atadMARehtsezilaitinI
eht,detcelessidohtemnoitazilaitinisihtretfaNOdenrutsirewopnehW
.deyalpsiddnadeveirtereblliwyromemhsalfniderotsdahuoytahtatad
On the MAIN OPS MENU, move the cursor to the "SYS-CND" item, and press the SET key.
1SERIAL OUTPUT NO PC-LINK SERIAL
2COMM.SET
(TO NEXT SUB-MENU)
3
COMPUTER LINK
(TO NEXT SUB-MENU)
4INITIALIZATION ALL-INIT INIT-RAM
5
SIMPLE MENUSTD
EXEC
[SYSTEM COND]
REFERENCE IMG
SYSTEM I/O
OBJECT TYPE COND
■■
■
■■■■■■■■■■■□□□□□
INITIALIZING