Technical data
Rev.February 1996 Conti-Dryer Page 29
CD 25-600
Temperature controller defective Replace
Thermistor defective Replace
Control PCB etc. defective Replace
B. Signal: Green lamp lights up + green diode on temperature controller
Cause Remidial action
The temperature controller does not
activate the heater
1. Circuit breaker overload for heater Reset QFM2. At recurrences call for a
cut out repairman.
2. Temperature controller defective Replace
3. Heater defective Replace
C: Signal: Green lamp lights up + temperature controller no light
Cause Remidial action
Temperature controller defective Replace
D: Signal: Temperature lower than temperature adjusted on temperature controller
Cause Remidial action
PT100 sensor defective Replace
Temperature controller defective Replace
Heater defective Replace