Operation Manual

10
E
Calibration and Measurement
Select a color mode for “Calibration” or “Measurement”, congure the colorimetric sensor settings, and perform “Calibration” or
“Measurement”.
Install a software for the colorimetric sensor in advance.
1. Connect a monitor to be adjusted to the computer with a USB cable.
2. Double-click the shortcut icon on the desktop.
3. Click “Calibration” or “Measurement” on the home screen.
4. Select a preset or user you want to perform “Calibration” or “Measurement” on.
User setting cannot be checked unless target values are set.
In the case of “Calibration”, each check takes approximately 30 to 80 minutes.
In the case of “Measurement”, each check takes approximately 1 to 4 minutes.
“COLOR MODE selection” screen for calibration
(“COLOR MODE selection” screen for measurement
in the case of measurement)
(2) (3)
(6) (5)
(4)
(1)
(1) Select target for “Calibration” or “Measurement”
(2) (
) Set/Change “User setting name (users of the color mode)” of the monitor
(A user-dened name must be set in 1 to 12 single-byte alphanumeric characters and symbols)
(3) Edit a target value le (when a target value is a le name) or create a new one (when target values are to be
newly created) (See pages 8 and 9.)
(4) (
) Select or ( ) deselect a target value le
(5) Time elapsed for last calibration
(6) Date and time of last calibration
5. Click “Next”.
Proceed to initialization and installtion of the colorimetric sensor.
Caution
Before performing “Measurement”, turn on the monitor to be measured and wait for at least 30 minutes until the luminance
stabilizes.
To perform more precise measurement:
- Set the “COLOR MODE” in the PICTURE menu of the monitor settings to the “COLOR MODE” to be measured.
- After the above conguration, wait for at least 30 minutes before performing measurement.
“Calibration” can be performed right after turning on a monitor to be calibrated, since it includes time to stabilize the
luminance.
Perform “Calibration” or “Measurement” within the operating temperature range of colorimetric sensor.
For the operating temperature range of colorimetric sensor, refer to the operation manual of colorimetric sensor.
During “Calibration” or “Measurement”, make sure that the ambient temperature does not change drastically.
If more than approximately 10°C of temperature change occurs inside the colorimetric sensor due to change of the ambient
temperature, an error will result.