4.3 Level Adj ustment
For correct weighing operations the scale should leveled prior to use. Turn the four
adjustable legs so that the bubble in the level gauge is positioned inside the center
circle. Then secure the locking nut.
When the Level Bubble is to the RIGHT.
Turn the Left Side Adjustable Legs Clockwise, or
turn the Right Side Adjustable Legs
Counterclockwise.
When the Level Bubble is to the LEFT.
Turn the Right Side Adjustable Legs Clockwise, or
turn the Left Side Adjustable Legs Counterclockwise.
When the Level Bubble is to the BACK.
Turn the Front Side Adjustable Legs C lockwise, or
turn the Rear Adjustable Legs Counterclockwise.
When the Level Bubble is to the FRONT
.
Turn the Rear Adjustable Legs Clockwise, or turn the
Front Side Adjustable Legs Counterclockwise.