Specifications

SG Series Programming SCPI Command Operation
M550129-03 Rev G 6-15
Command Description
SCPI
in millivolts.
:PROTection?
Returns the set overvoltage protection trip point in volts
(default) or in millivolts.
C
[:LEVel] <NRf>
Sets the overvoltage protection trip point in volts (default) or
in millivolts.
C
[:LEVel?]
Returns the set overvoltage protection trip point in volts or in
millivolts.
C
:TRIPped?
Returns 1 (TRIPPED) or 0 (UNTRIPPED) state of the
overvoltage protection CIRCUIT.
C
:STATe?
Returns the state 1 (ON) or 0 (OFF) If the overvoltage
protection is enabled.
C
:CLEar
Clears the overvoltage protection CIRCUIT. C
:RAMP <NRf> <NRf>
Sets the output voltage to ramp from the present value to the
specified value (first argument) in the specified time (second
argument). See Ramp Function description Section 6.5.3.
N
:RAMP?
Returns 1 if the ramp is in progress, and 0 if the ramp is
completed.
N
:ABORt
Aborts ramping and clears trigger mode. N
:ALL?
Returns the ramping status of all channels. N
:HTRIGgered <NRf>
Sets the value of the output voltage ramp to be implemented
when the hardware trigger is received.
N
:HTRIGgered?
Returns the value of the output voltage ramp to be
implemented when the hardware trigger is received.
N