Operating instructions
2
Contents
1.0 Introduction .........................................................................................................5
1.1 Intended Use .........................................................................................................................................5
1.2 SRPM Function .....................................................................................................................................5
2.0 Parts List ...............................................................................................................6
1.1 SRPM Parts List .....................................................................................................................................6
1.2 Model SRAN (Option) Parts List ......................................................................................................6
1.3 Accessories (Option) Parts List .......................................................................................................6
Installation Instructions ....................................................................7
3.0 Mounting and Wiring ........................................................................................9
3.1 Mounting ..............................................................................................................................................9
3.2 Wiring Electrical box (rough in) ....................................................................................................9
3.3 Attaching Pressure Tubing ............................................................................................................10
4.0 Model SRAN Room Annunciator Installation ............................................. 11
4.1 Alarm Relay Output ..........................................................................................................................11
4.1.1 Model SRAN Remote Annunciator Wiring ..................................................................11
4.1.2 Non-Setra Remote Annunciator .....................................................................................11
4.2 Audible Alarm .....................................................................................................................................12
5.0 Door Status Switch Wiring ( see Figure 3, J) ............................................... 12
5.1 Wiring 12
6.0 Electrical Installation ....................................................................................... 13
6.1 Analog Output (J5) ...........................................................................................................................13
User’s Operating Instructions ....................................................... 15
7.0 Startup and Operation ................................................................................................ 17
7.1 Power-up ..............................................................................................................................................17
7.3 SETUP UNIT SCREEN ........................................................................................ 19
7.3.1 Setup Unit Operation .........................................................................................................19
7.3.2 Entering Data .........................................................................................................................19
7.3.2.1 Password Protection .................................................................................................19
7.3.2.2 Display Averaging .....................................................................................................19
7.3.3 Data Entry Screen .................................................................................................................19
7.4 SETUP ROOM SCREEN ..................................................................................... 20
7.4.1 Setup Room Operation ......................................................................................................20
7.4.2 Entering Data .........................................................................................................................20
7.4.3 Data Entry Screen .................................................................................................................20
27
159 Swanson Road, Boxborough, MA 01719
Tel. 800-257-3872; Fax 978-264-0292; Web:www.setra.com
12.0 Electrical
Power
1) Model SRPM-XXXXX-V1-X
For the high voltage model the input voltage range is from 85
VAC to 265 VAC, 50 Hz or 60 Hz
All the internal supply voltages are generated using the Switching
Mode Power Supply (SMPS).
All the internal supply voltages and their ratings as follows:
Internal Supply Voltage Max. Current Consumption
+5 V 800 mA
+3.3 V 200 mA
+15 V 200 mA
2.) Model SRPM-XXXXX-A1-X
For the lower voltage model the input voltage range is from 18 to
32 VAC, 50 Hz or 60 Hz
All the internal supply voltages and their ratings as follows:
Internal Supply Voltage Current Consumption
+5 V 800 mA
+3.3 V 200 mA
+15 V 200 mA
Door Status Switch SPDT or SPST NO