Specifications
20
Appendix II
Calibration
Your Setra balance is highly stable, and requires infrequent calibration. While it may be
calibrated as often as desired, periodic calibration is adequate for most applications.
FULL-SCALE CALIBRATION
To calibrate your balance with a full scale mass use the following procedure:
1. Press the clear key.
2. Press the tare key.
3. Place the mass on the pan.
4. Press the cal key.
CALIBRATION WITH A SPECIFIC MASS
To calibrate your balance with a specific known mass (but less than full scale), use the
following procedure:
1. Press the clear key.
2. Press the tare key.
3. Press the conv key until the balance reads in the units of the known mass.
4. Place the known mass on the pan.
5. Enter the value of the known mass.
6. Press the cal key.
NOTE: To ensure maximum accuracy, a mass should be used which is as close
as possible to the full-scale value of the balance.