Specifications

11
REDUCED RESOLUTION WEIGHING
In the presence of severe vibration or draft a stable display can often be achieved by
reducing the displayed resolution.
1. Press the x10 key to reduce displayed reading by one digit.
2. To return to the normal displayed resolution, press the x10 key again.
PERCENT DEVIATION
To calculate the amount by which a weight varies from a reference, follow this
procedure:
1. Press tare.
2. Place the reference weight on the pan.
3. Press the % key. After acquiring a stable reading, the display will read “100.000
pc”.
4. Press the tare key. After acquiring a stable reading, the display will read
“0.000pc”. The display shows percent deviation.
5. Remove the reference weight.
6. Place the weight to be measured on the pan.
7. Wait for the “OK” symbol to appear. Read the display. The display indicates
percent deviation from the reference.
8. Remove the weight.
9. Repeat Steps 6-8 as many times as desired.
10. Press clear to return to weighing.
FILLING TO A TARGET WEIGHT - FORMULATION BY PERCENTAGE
It is often useful to measure out a certain weight of substance which is a mixture (by
percent) of component parts. This is easily done by using the percent function.
1. Place a container on the pan.
2. Press tare.
3. Press the key.
4. Enter the target weight.