Results of Compliance Testing for EMC Directives
Products: Top Loading Scales
Model EZ2-500
Model EZ6-500
Model EZ2-2000
Model EZ6-2000
Model EZ2-5000
Model EZ6-5000
EMC Directive 1994 EN 50081-1 and EN 50081-2 Electromagnetic Compat-
ibility Generic Emission Standard for Industrial, Scientific and Medical Equip-
ment
Test No. Level Criteria Results
EN55011 Radiated RF 30-230 MHz < 30 dB uV/m *Pass
EN55011 Radiated RF 230-1000 MHz < 37 dB uV/m *Pass
EMC Directive EN 50082-1 1995 Electromagnetic Compatibility
Generic Immunity Standard for Residential, Commercial, Light Industry
Test No.: IEC801-2 ESD Level: 8kV Air Discharge
Criteria: B - Must Resume Normal Operation After Test Results: Pass
Test No.: IEC801-2 ESD Level: 4kV Direct Discharge
Criteria: B - Must Resume Normal Operation After Test Results: Pass
Test No.: IEC801-3 Radiated RF Level: 3V/m 27-500 MHz
Criteria: A - Must Operate As Specified Results: **Pass
Test No.: IEC801-4 Transient Burst Level: 5 ms Burst, 5KHz
Criteria: B - Must Resume Normal Operation After Test Results: Pass
* Use of ferrite bead on power lead required.
** Typical susceptibility to radiated frequencies is < 0.0005% FS. At certain
discreet frequencies there is the potential for a maximum effect of 0.18% FS.