Specifications
26
5. Press the zero key. The scale will now perform the normal wake up
routine.
If the scale displays any unusual messages during the warm-up period, or
does not return to normal operation, contact an Authorized Setra Service
Center or Setra Systems, Inc., as service may be required.
If the scale displays the “bUSY” message for an extended period of time, or
the displayed reading is unstable, too much vibration or draft may be present.
Relocate the scale away from the source of vibration or shield the scale from
draft in order to correct this condition. If vibration or draft is not the cause of
the instability, service may be required.
If the scale displays “d-noS”, it may have been subjected to an extreme
discharge of static electricity or power line fluctuations. The scale can be
easily restored to operation by following the procedures outlined in steps
#4 and #5 above. Note, the user setups will be restored to the original
factory defaults.
If the RS-232 interface does not function correctly, first make certain that
you are using the correct RS-232 cable, and that it is securely fastened to
both the scale and the peripheral device. If data transmission or reception is
still not possible, be certain that both the scale and peripheral device serial
ports are set up correctly (baud rate, parity, handshaking, etc.). See Appen-
dix IV.
When communicating with certain types of devices, especially computers, it
may be necessary to “cross” the receive and transmit lines of the interface.
That is, the receive line of the scale must connect to the transmit line of the
external device. Likewise, the transmit line of the scale must connect to the
receive line of the external device. Special cables can be purchased for this
purpose from a computer dealer.
If the problem still persists, contact your local authorized Setra dealer, refer
to the Setra Easy Count Technician’s Guide or consult Appendix IX, “Limited
Two-Year Scale Warranty,” for more information.
To obtain a copy of the Setra Easy Count Technician’s Guide contact your
local authorized Setra dealer.