Specifications
25
Appendix IV
In Case of Difficulty
Setra Easy Count scales are designed to be easy to use. If, however, you have
reached a state in which the scale will no longer follow your instructions,
you may wish to “reset” the scale. It is possible to do so by following this
sequence.
1. Remove all weight from the pan.
2. Press the clear key (EZ6) or the zero key (EZ2).
If this procedure does not solve the problem, try briefly unplugging the unit
from its power source, then plug the unit in again.
To print a copy of the scale’s current setup information use the following
procedure. Connect the scale to a printer or computer via the RS-232 serial
port. Power up the scale and then hold the count key down while the scale
completes the countdown and displays the scale capacity, i.e., -2-, -1-, C
xxxxg. The scale will display “rEF n.n”, the software version while printing
the current setup information. Press the clear key (EZ6) or the zero key
(EZ2) twice to return to the normal weighing mode.
If the scale does not return to normal operation, you may want to com-
pletely reset the scale’s memory. This will cause all of the user setups to
return to the factory defaults and WILL ERASE ANY SPAN CALIBRATION that
may have been performed. To reset the scale’s memory use the following
procedure:
1. Remove power from the scale by unplugging the AC adaptor.
2. Hold down the zero and count keys simultaneously.
3. While holding these two keys down, restore power to the scale, and
continue to press the keys until “d-noS” is displayed. This puts the
scale in the diagnostic mode.
4. Press the zero key. The scale will now perform tests to verify its opera-
tion and reset all of the setups to the factory defaults. The scale will
then display “PASS”.