Specifications

20
Appendix II
Calibration
Your Setra Easy Count scale is highly stable, and requires infrequent calibra-
tion. While it may be span calibrated as often as desired, the factory calibra-
tion is adequate for most counting applications. For precise weighing appli-
cations more periodic calibrations may be desirable. To perform a span
calibration on an Easy Count scale, use the following procedure:
1. Remove all objects from the pan.
2. Make certain the scale is level.
3. Press the zero key to make sure the scale is reading “0.00”.
4. Press and hold the zero key and then press the count key.
5. The scale will display “SPAn”.
6. Press the zero key to enter the span menu and display “L xxxxx”,
where xxxxx is the scale’s first calibration point.
7. Press the count key repeatedly to display each calibration point avail-
able.
To ensure maximum accuracy the calibration point selected should be
as close as possible to the scale’s full capacity.
8. When the desired calibration point is displayed, load the displayed
weight on the pan, and press the zero key.
9. The scale will display “bUSY”, and then return to the normal weighing
mode.
10. Remove the weights from the pan, press zero to zero the display.