Specifications

8
CALIBRATION
To perform a span calibration, use the following procedure:
STEP 1.
MODE
TARE
POWER
0.00
0.00
Press the TARE key to zero the balance.
STEP 2.
TARE
POWER
MODE
SPAn
0.00
Hold down the TARE key then press the MODE key. Release both keys, the
display will read “SPAn” for span calibration.
STEP 3.
2000SPAn
MODE
TARE
POWER
Press the MODE key to enter the span calibration menu. The balance’s full
capacity will be displayed. You can calibrate the balance at full capacity or at
a lower calibration point. To view the set calibration points, press the TARE
key until the desired weight is displayed.
To restore the factory calibration, press the TARE key to display “UnSPAn”
then press the MODE key. To escape, press the TARE key to display “ESCAPE”
then press the MODE key.
STEP 4.
MODE
TARE
POWER
ACAL
2000
Place on the pan calibration weight(s) equal to the displayed weight and
press the MODE key. The balance will display “ACAL” and then return to the
normal weighing mode. If the display reads “nOCAL”, check the weights and
weighing environment. This calibration procedure can only correct for ±1%
span shift.