Specifications

16
For slow response
Perform Step 4., then press the TARE key once to display “SLO”, then
press the MODE key to select the slow response rate. The balance will
return to the normal weighing mode.
For auto response
Perform Step 4., then press the TARE key twice to display “AUTO”, then
press the MODE key to select the auto response rate. The balance will
return to the normal weighing mode.
TEMPERATURE COMPENSATION CALIBRATION
If your balance is frequently moved from one location to another or if the
room temperature varies more than a few degrees during the day, the
balance may display a slight temperature induced offset. Your Setra balance
is capable of compensating for this offset so that varying temperatures will
not affect its performance.
If your balance is used in different locations, leave it where the temperature
is either higher or lower than the operating environment for at least four
hours. Otherwise, perform this calibration first thing in the morning or last
thing in the evening to expose the balance to the largest possible tempera-
ture fluctuation. Then, place the balance in its operating environment and
run this calibration procedure. Over the next three hours it will monitor any
weight fluctuations that occur and record these temperature induced varia-
tions into memory. During this calibration period the balance cannot be
used and must not be disturbed. Once this calibration procedure is com-
pleted, the balance will be able to internally compensate for the offset.
To perform the temperature compensation calibration:
STEP 1.
Leave the balance for at least four hours in the initial environment as ex-
plained above. The balance does not have to be powered on.
STEP 2.
Move the balance to its operating environment. Unplug the balance (if
powered on) for 10 seconds and then power on. The balance will count
down and should display 0.00.
Note: The 0.00 reading may not be very stable if the balance is warming up
or cooling down.