Specifications
13
Note: To escape anytime during this procedure, press the TARE key until
"ESCAPE" is displayed and press the MODE key.
STEP 3.
rEF2.00rEF
MODE
TARE
POWER
Press the MODE key to display the software reference number.
STEP 4.
MODE
TARE
POWER
0.00
rEF2.00
Press the TARE key to return to the normal weighing mode.
RESTORING THE FACTORY DEFAULT SETUPS
The many features described in this section allow the user to customize the
balance to suit a particular application. However, in doing this it is also
possible to inadvertently set up the balance in such a way that it does not
operate as expected. To reset the factory defaults to: MODE key will print a
stable reading and then advance the printer to the next line, auto response
rate, 2400 baud, no parity, follow the steps below.
Note: Restoring the factory defaults will return your balance to all of the
factory span and temperature calibration settings. You MUST
recalibrate (span) your balance after restoring the factory defaults. If
you are experiencing a temperature induced offset, you should also
run the temperature compensation calibration procedure.
STEP 1.
TARE
POWER
MODE
SPAn
0.00
Hold down the TARE key then press the MODE key. Release both keys, the
display will read "SPAn".